ZOLL Pro-padz Sterile Multi-Function Electrodes (MFE) Part Number: 8900-4012 Product Usage: These MFE Electrodes are use
FDA device recall Z-0052-2015 · posted 2014-10-09 · Terminated
Recall details
- Recalling firm
- Bio-Detek, Inc.
- Reason for recall
- Product sterility is not assured
- Action taken
- Bio-Detek notiifed customers of the expanded recall on September 16, 2014 for Lot 1914 and September 22, 2014 for Lots 1614, 1814, 2114 via UPS overnight delivery. Customers were requested to return the suspect product(s) in their possession. A Device Correction Response Return Form was provided to each Customer. For questions call 1-800-348-9011 or 1-978-421-9460.
- Root cause
- Material/Component Contamination
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 237 cases (6/case)
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Australia, Canada, Belgium, France, and Germany.
- Date initiated
- 2014-09-16
- Date posted
- 2014-10-09
- Date terminated
- 2016-05-03
- Location
- Pawtucket, RI