ZOLL Pro-padz Sterile Adult Multi-Function Electrodes with 10-ft (3 m) leadwires Part number: 8900-4052 Product Usage: T

FDA device recall Z-0054-2015 · posted 2014-10-09 · Terminated

Recall details

Recalling firm
Bio-Detek, Inc.
Reason for recall
Product sterility is not assured
Action taken
Bio-Detek notiifed customers of the expanded recall on September 16, 2014 for Lot 1914 and September 22, 2014 for Lots 1614, 1814, 2114 via UPS overnight delivery. Customers were requested to return the suspect product(s) in their possession. A Device Correction Response Return Form was provided to each Customer. For questions call 1-800-348-9011 or 1-978-421-9460.
Root cause
Material/Component Contamination
Status
Terminated
Product code
MKJ
Quantity affected
193 cases (6/case)
Distribution
Worldwide Distribution - US Nationwide and the countries of: Australia, Canada, Belgium, France, and Germany.
Date initiated
2014-09-16
Date posted
2014-10-09
Date terminated
2016-05-03
Location
Pawtucket, RI