Atlas FDA

Zoll pedi.padz Reduced Energy Multi-Function Electrodes (MFE) Part number: 8900-0401 Used in connection with Zoll M Seri

FDA device recall Z-1011-2012 · posted 2012-02-13 · Terminated

Recall details

Recalling firm
Bio-Detek, Inc.
Reason for recall
Lack of an adequate silicone seal around the periphery of the attenuator enclosure may allow fluid to enter.
Action taken
Zoll Medical issued Urgent Device Correction notices dated 12/30/11 via UPS/Registered Mail. The letter described the issue and identified the affected product. ZOLL recommended that customers alert their staff of the issue, that the electrodes should be kept away from a fluid environment, to use caution in handling the electrodes if they are subjected to a fluid environment and maintain safe defibrillation practice, and seek a replacement if available. Also customers should contact Zoll 1-800-348-9011 for product replacement. The firm plans to replace the affected lots free of charge.
Root cause
Process control
Status
Terminated
Quantity affected
3,080 units
Distribution
Worldwide Distribution -- US and Canada.
Date initiated
2011-12-30
Date posted
2012-02-13
Date terminated
2012-12-11
Location
Pawtucket, RI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ZOLL PEDIPADZ REDUCED ENERGY ELECTRODE (K072430)ZOLL Medical Corporation2007-11-06