Atlas FDA

Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 9 mm diameter, 130 mm length, non-sterile, catalog no. 43

FDA device recall Z-1272-06 · posted 2006-07-27 · Terminated

Recall details

Recalling firm
Zimmer Inc.
Reason for recall
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
Action taken
U.S. and international distributors were notified of the recall via letter dated 6/30/2006 to check customer inventories and to return recalled product to the firm. Trial implants were redesigned and reworked to eliminate the problem.
Root cause
Other
Status
Terminated
Product code
HWT
Quantity affected
324 total
Distribution
Nationwide and worldwide including Australia, Canada, Germany, Italy and Spain.
Date initiated
2006-06-16
Date posted
2006-07-27
Date terminated
2007-12-05
Location
Warsaw, IN

Monitor recalls programmatically