Zimmer Trabecular Metal Reverse Shoulder System Instrumentation Conical Reamer, instrument for shoulder arthroplasty, RE
FDA device recall Z-0675-2007 · posted 2007-03-29 · Terminated
Recall details
- Recalling firm
- Zimmer Inc.
- Reason for recall
- The reamer may lock up on the threads of the reamer body rather than spin freely as intended by design.
- Action taken
- Zimmer sales staff was notified by letter dated 2/1/07 to pull any of the product from stock and to provide the name and address of their hospital customer accounts for further notification by Zimmer.
- Root cause
- Other
- Status
- Terminated
- Product code
- LXH
- Quantity affected
- 4
- Distribution
- California, Ohio and Texas.
- Date initiated
- 2007-02-01
- Date posted
- 2007-03-29
- Date terminated
- 2007-05-04
- Location
- Warsaw, IN