Atlas FDA

Zimmer Plates and Screws (ZPS) 4.0 mm Cancellous Screw Fully Threaded, 12 mm Length 47484001200 4.0 mm Cancellous Screw

FDA device recall Z-1677-2016 · posted 2016-05-16 · Terminated

Recall details

Recalling firm
Zimmer Manufacturing B.V.
Reason for recall
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
Action taken
All distributors will be notified via electronic mail. Distributors with affected lots will also be notified via courier. Hospitals risk managers and surgeons will be notified via recall notification letter, dated May 4, 2016.
Root cause
Packaging change control
Status
Terminated
Product code
HWC
Quantity affected
3955
Distribution
Worldwide distribution. US nationwide, Korea, Japan, Argentina, Brazil, Singapore, Taiwan, China, El Salvador, Australia, India, Hong Kong, Chile, Germany, Italy, France, Spain, UK, Sweden, Canada.
Date initiated
2016-05-04
Date posted
2016-05-16
Date terminated
2017-07-05
Location
Mercedita, PR

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ZIMMER PLATES AND SCREWS SYSTEM SCREWS ONLY (K112885)Zimmer, Inc.2011-12-19