Atlas FDA

Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 46 mm dia. articular surface, large, ste

FDA device recall Z-0028-06 · posted 2005-10-07 · Terminated

Recall details

Recalling firm
Zimmer Inc.
Reason for recall
The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement.
Action taken
Consignees were notified via letter dated August 23, 2005 or by email.
Root cause
Other
Status
Terminated
Product code
KWS
Distribution
Nationwide, Australia, Canada, England, Germany and Italy.
Date initiated
2005-08-23
Date posted
2005-10-07
Date terminated
2005-12-08
Location
Warsaw, IN