XSYSTEMS Dilution Buffer, list number 9519-02; for In Vitro diagnostic use, Bovine gamma globulin in Phosphate buffer, w
FDA device recall Z-1474-04 · posted 2004-09-18 · Terminated
Recall details
- Recalling firm
- Abbott Laboratories HPD/ADD/GPRD
- Reason for recall
- The XSYSTEMS Dilution Buffer when used with TDx/TDxFLx Benzodiazepines assay can cause shifts in control values and patient results.
- Action taken
- Abbott sent Device Correction letters dated 9/1/04 to all customers receiving the TDx/TDxFLx Benzodiazepines assay, list 9674, informing them of the shift in control values and patient results when used with the XSYSTEMS Dilution Buffer, list 9519-02. The letter gave examples of the effects observed in the benzodiazepines assay performance, and listed the following actions to be taken until Abbott completes their investigation: * During the use of a single lot of buffer, recalibrate the TDx/TDxFLx Benzodiazepines assay every 14 days. * Perform a new calibration run on the TDx/TDxFLx Benzodiazepines assay when a new lot number of XSYSTEMS Dilution Buffer is used. * These actions are in addition to the package insert calibration and quality control requirements. * Follow your laboratory procedures for notification of the health care providers that you serve. Any questions were directed to Abbott''s Customer Service at 1-877-4ABBOTT.
- Root cause
- Other
- Status
- Terminated
- Product code
- JJE
- Quantity affected
- 22,442 cartons
- Distribution
- Nationwide and internationally to Argentina, Australia, Brazil, Chile, Colombia, Costa Rica, El Salvador, Germany, Guatemala, Honduras, Hong Kong, Mexico, Panama, Peru, Puerto Rico, Singapore, South K
- Date initiated
- 2004-09-01
- Date posted
- 2004-09-18
- Date terminated
- 2004-11-30
- Location
- Abbott Park, IL