XiO Radiation Treatment Planning System, XiO Release 1.5.0 and above. Used to create treatment plans for any cancer pati

FDA device recall Z-2574-2011 · posted 2011-06-15 · Terminated

Recall details

Recalling firm
Computerized Medical Systems Inc
Reason for recall
In XiO, when Telepathy is exited, re-entered, and the temporary plan re-called, the OLD beam weight will be displayed but the dose distribution will reflect the modified beam weight.
Action taken
Computerized Medical Systems, Inc. sent an CUSTOMER ADVISORY letter to all affected customers in May 2008. The letter identified the product, the problem and the action needed to be taken by the customer. The issue was resolved in XiO Release 4.50.00 which has been available since June 2009. A User Notice was sent in March 2011, to all the customers reminding them of the availability of the correction. This notice was sent hardcopy via US Mail but there was no return receipt. For further questions, please call ( 408) 380-8023
Root cause
Software design
Status
Terminated
Product code
muj
Quantity affected
936
Distribution
Worldwide Distribution -- USA (nationwide) including Puerto Rico, and countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Belgium, Bolivia, Bosnia and Herzegovina, Brazi
Date initiated
2008-05-14
Date posted
2011-06-15
Date terminated
2014-04-08
Location
Maryland Heights, MO