X-Sizer Thrombectomy Catheter System (ev3) Model number XD-CS1150-45 (for US distribution). Model number XR-CS1150-45 (f

FDA device recall Z-1107-06 · posted 2006-06-14 · Terminated

Recall details

Recalling firm
Ev3, Inc
Reason for recall
Loss of Operating Vacuum during use-It has been determined that a small percentage of X-Sizer Thrombectomy Catheter Systems from specific lots may lose operating vacuum prematurely due to a system leak within the Control Module. This failure can lead to the inaility to complete the procedure of thrombus removal and the necessity to remove the device prior to achieving a satisfactory result.
Action taken
A medical Device Recall Letter dated 02/13/06 was sent to hospitals along the affected device lot numbers the hospital should have received on a Device Reconciliation Form. ev3 is asking for affected devices to be returned. Returned devices will be replaced at no cost.
Root cause
Other
Status
Terminated
Product code
MCW
Quantity affected
Model number XD-CS1150-45 (US) 534 devices. Model number XR-CS1150-45 (OUS) 504 devices.
Distribution
World wide-AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MN, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. OUS to: Argen
Date initiated
2006-02-14
Date posted
2006-06-14
Date terminated
2006-12-26
Location
Plymouth, MN