X-Ray Diagnostic Table, Model No. YSF-300 Use in combination with an X-ray high voltage generator, an X-ray tube unit, a

FDA device recall Z-1585-2016 · posted 2016-04-28 · Terminated

Recall details

Recalling firm
Shimadzu Medical Systems
Reason for recall
Report of unintentional movement of table to reverse tilting.
Action taken
Shimadzu sent an Urgent Voluntary Medical Device Recall Notice to all affected customers on April 8, 2016. Shimadzu Corporation has identified a potential issue in the Shimadzu X-ray Diagnostic Table. The letter informs the customers of the problems identified and the actions to be taken. Customers with questions or concerns are instructed to contacted Shimadzu National Technical Support at (800) 228-1429. Shimadzu will send out an updated customer notification letter to inform customers that the liquid penetration may cause an electrical short circuit, causing the table to unintentionally tilt in reverse. If this unintentional reverse tilting occurs, there is a potential risk of injury to the patient and/or the user. For questions regarding this recall call 310-217-8855, ext 139.
Root cause
Device Design
Status
Terminated
Product code
IXR
Quantity affected
185 units
Distribution
Nationwide Distribution
Date initiated
2016-03-31
Date posted
2016-04-28
Date terminated
2017-01-02
Location
Torrance, CA