Atlas FDA

Wound Drain - Label reads in part "Elutia ***Hubless Coated Silicone Surgical Wound Drain***" Two different sizes, 10 un

FDA device recall Z-1680-2014 · posted 2014-06-02 · Terminated

Recall details

Recalling firm
Bacterin International, Inc.
Reason for recall
The recalling firm reported there was inadvertent omission of quarterly (Q4 - 2013 and Q1 - 2014) dose audits/verifications on the affected products. All of the affected products were appropriately sterilized
Action taken
Bacterin sent e-mail and FedEx letter with return receipt request to it's single consignee on May 15, 2014. Consignee is asked to separate the recalled product in a secure location for return to Bacterin International, Inc. Consignee is asked to complete tracking/verification form included in the letter/e-mail and call 1-888-886-9354 from 8:00AM to 5:00PM (Mountain Time) or send an e-mail to CS@Bacterin.com to obtain a Return Authorization (RA) number.
Root cause
Use error
Status
Terminated
Product code
OEI
Quantity affected
390
Distribution
Distributed to one foreign consignee in Lebanon.
Date initiated
2014-05-15
Date posted
2014-06-02
Date terminated
2014-06-02
Location
Belgrade, MT

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ELUTIA COATED CLOSED SURGICAL WOUND DRAIN (K063245)Bacterin International, Inc.2007-11-28