Atlas FDA

Waterlase MD, dental laser unit, Model #6200218, serial numbers 10500210 to 105202680, inclusive.

FDA device recall Z-0543-05 · posted 2005-02-15 · Terminated

Recall details

Recalling firm
Biolase Technology Inc
Reason for recall
Reports that the Waterlase MD, a dental laser unit, water bottle broke under pressure during installation.
Action taken
Water bottles will be replaced by trained Biolase representatives. During that visit, the representative will verify that the bottle pressure is set correctly and the existing bottle is replaced with a new bottle.
Root cause
Other
Status
Terminated
Product code
LYB
Quantity affected
206 units
Distribution
Nationwide, Canada, Germany, Ireland, France, United Arab Emirates, Lebanon, Greece, Hong Kong, Argentina, Indonesia, Estonia, Spain, Taiwan
Date initiated
2005-01-17
Date posted
2005-02-15
Date terminated
2005-08-17
Location
San Clemente, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
WATERLASE (K031140)Biolase Technology, Inc.2004-07-07