WarmAir Model 135 Convective Warming Device, Cincinnati Sub-Zero Products, Inc. Used to prevent hypothermia and/or reduc

FDA device recall Z-2888-2011 · posted 2011-07-27 · Terminated

Recall details

Recalling firm
Cincinnati Sub-Zero Products Inc
Reason for recall
Due to its design the circuit breaker on the device could trip causing it to shut down and not provide heated air to the blanket.
Action taken
Cincinnati Sub-Zero Products, Inc. sent an 'URGENT Medical Device Recall Field Action Notice" dated July 12, 2011 to all affected customers. The letter includes: description of product, problem, and information to request a repair tubing kit free of charge. It also asks customers to complete and return a "Recall Response Form". For additional information e-mail WA135-2011@cszinc.com or call (800) 989-7373 or (513) 772-8810.
Root cause
Device Design
Status
Terminated
Product code
DWJ
Quantity affected
5474 units
Distribution
Worldwide Distribution: (USA) Nationwide, and the countries of Australia, Brazil, Bulgaria, Canada, Chile, Ecuador, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Malaysia, Mexic
Date initiated
2011-06-01
Date posted
2011-07-27
Date terminated
2013-08-02
Location
Cincinnati, OH