Atlas FDA

Warm Air Model 135 Warming Units.

FDA device recall Z-0374-05 · posted 2004-12-30 · Terminated

Recall details

Recalling firm
Cincinnati Sub-Zero Products
Reason for recall
Burning smell, burned microprocessor board. The firm is modifying the connectors and wiring gauge configuration of the device.
Action taken
The firm notified consignees by certified letter on November 5, 2004.
Root cause
Other
Status
Terminated
Product code
BTE
Quantity affected
649 units.
Distribution
The product was distributed throughout the United States and to foreign countries including Saudi Arabia, Taiwan, Japan, and Canada.
Date initiated
2004-11-05
Date posted
2004-12-30
Date terminated
2008-10-29
Location
Cincinnati, OH