Wallach LL100 Cryosurgical Freezer No Gas Fitting CO2 System Model: 900019x.
FDA device recall Z-0095-2010 · posted 2009-10-29 · Terminated
Recall details
- Recalling firm
- Cooper Surgical, Inc.
- Reason for recall
- CO2 Cryosystem thermal load may exceed the exhaust gas phase, resulting in ice formation in the exhaust valve. Should this occur, the Cryo tip may cool at a less than optimal rate, resulting in ice formation in the exhaust valve.
- Action taken
- Cooper Surgical notified customers by letter titled "Notice of Recall - Urgent" dated September 29, 2009 advising them of the problem and requesting return of the device. International accounts were notified by letter Beginning October 1, 2009. For further information, contact Cooper Surgical by fax at 1-203-799-2002 or phone at 1-203-799-2000.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- GEH
- Distribution
- Worldwide Distribution -- United States, Canada, Australia, Belgium, England, France, Germany, Greece, Hungary, Iceland, Italy, Ireland, Latvia, Lithuania, Netherlands, Portugal, Spain, Solvenia, Swed
- Date initiated
- 2009-09-29
- Date posted
- 2009-10-29
- Date terminated
- 2012-04-17
- Location
- Trumbull, CT