WalkAide Stimulator System Control Module, Part Number 20-0100; component of the WalkAide Stimulator System; distributed

FDA device recall Z-1435-2009 · posted 2009-06-03 · Terminated

Recall details

Recalling firm
Innovative Neurotronics, Inc.
Reason for recall
Knob responsible for regulating the amount of stimulation administered by the WalkAide System control unit may fall off while device is in use.
Action taken
Innovative Neurotronics Inc notified the consignee via e-mail on August 6, 2008 of the affected product. The Consignee was asked to place a stop shipment on the WalkAides and to return inventory for replacement. Further questions can be directed to Innovative Neurotronics Inc at 1-512-721-1906.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
GZI
Quantity affected
137 units
Distribution
Nationwide Distribution to the state of GA.
Date initiated
2008-08-06
Date posted
2009-06-03
Date terminated
2009-07-28
Location
Austin, TX