VT-0101-50 Otological Ventilation Tubes, Catalog Number VT-0101-50, Donaldson, with tab, 1.14 mm Lumen, Silicone, Steril
FDA device recall Z-0348-2011 · posted 2010-11-18 · Terminated
Recall details
- Recalling firm
- Micromedics, Inc.
- Reason for recall
- Micromedics, Inc is conducting a product recall on a number of medical device products because of weak seals of the sterile pouches, potentially resulting in a non-sterile product which may cause transmission of disease or infection.
- Action taken
- Micromedics issued a Product Recall letter dated March 15, 2010 to customers, describing the product being recalled, the product sterility issue, and actions to be taken. The letter advised customers to contact Customer service to return any affected product to Micromedics for replacement. Customers can contact Micromedics at 1-800-624-5662 or 651-452-1977 ext. 226 concerning this recall.
- Root cause
- Packaging process control
- Status
- Terminated
- Product code
- ETD
- Quantity affected
- VT 213 (1982 pouches); ET 50 (500 pouches)
- Distribution
- Nationwide Distribution in the states of CA, CT, DE, FL, GA, ID, IL, IN, MD, MA, MI, MN, NJ, NY, NC, PA, TX, UT, and WI. FRANCE, NETHERLANDS, BELGIUM, UNITED KINGDOM, CANADA, ISRAEL, PERU, SPAIN, IREL
- Date initiated
- 2010-03-15
- Date posted
- 2010-11-18
- Date terminated
- 2011-06-21
- Location
- Eagan, MN