Atlas FDA

VS3 Iridium - Visionsense Infrared (IR) Fluorescence Miniature Microscope (MMS) - 785nm Product Number: 175-0012

FDA device recall Z-0281-2023 · posted 2022-11-23 · Open, Classified

Recall details

Recalling firm
Visionsense, Ltd.
Reason for recall
XXX
Action taken
Medtronic issued Urgent Medical Device Correction Letter on 17-OCTOBER-2022 to US consignees via UPS 2-day delivery. For outside U.S. countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and action to take: 1.Prior to use, visually check the Beam-Combiner and MMS optical fiber cable for signs of being pinched, kinkedor damaged. Avoid folding the cables and handle with care to prevent damage. 2.After the white balance has been performed as per IFU guidance, test the EleVision IR system at the beginningof the surgical procedure as follows: a)Turn on the laser. b)For the laparoscopic configuration of the EleVision IR system, point the distal end of the endoscopetowards a bright surface, keeping the scope approximately 3 cm from the target. This prevents thewhite light from scattering. For the open configuration of the EleVision IR system, position the distalface of the MMS approximately 30 cm from any object. If you see a message on the screen indicatingthat the device is too close to tissue, move the device further away from the object until this messagedisappears. c)After two (2) seconds, the monitor screen should show a message indicating that the laser is on, ablinking red dot, and a running clock. If all these indications are seen properly on the monitor screen,turn off the laser after five (5) seconds. If the screen does not show an indication that the laser is on, a blinking red dot, and a running clock, the laser is not functioning properly and the EleVision IR system should not be used in the surgical procedure. If any pinching, kinking, or damage of the optical fiber cable is noted, or if the system does not work properly during the above-described laser test, do not use the system. Please report a complaint and arrange for servicing of the Beam Combiner and / or MMS device by contacting Medtronic at 1-800-255-6774 or rs.serviceexperience@medtronic.com. Additional actions: "Immediately notify
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
OWN
Quantity affected
34 units: 15 units OUS; 19 units US
Distribution
US Nationwide distribution.
Date initiated
2022-10-17
Date posted
2022-11-23
Location
Tikva, Israel

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
VS3-IR-MMS System (K150018)Visionsense, Ltd.2015-06-17