Atlas FDA

Vortex Surgical I2 Injection Kit, VS0500

FDA device recall Z-1252-2026 · posted 2026-02-04 · Open, Classified

Recall details

Recalling firm
Vortex Surgical Inc.
Reason for recall
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Action taken
On December 16, 2025, the firm began notifying customers via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were instructed to discontinue use of affected product and to return product for Vortex Surgical to evaluate the seals; use UPS account 23623Y to ship unused products to Vortex Surgical Inc for replacement or refund at: Attn: Andrew Ritts Recall Tyvek Seal, Vortex Surgical Inc., 4 Research Park Dr., Suite 124, St. Charles, MO 63304; complete and return the Acknowledgement and Receipt Form via mail or email to: ARitts@VortexSurgical.com. Actions have been implemented to improve training of individuals associated with sealing and inspecting seals to prevent reoccurrence. The sealer associated with the issue is being updated to include a guard to prevent reoccurrence. If you have any questions, contact the Director of Regulatory Affairs at 635-778-4350 or email to: ARitts@VortexSurgical.com from 8am to 5pm.
Root cause
Employee error
Status
Open, Classified
Product code
KDD
Quantity affected
14,789 (8651 US; 6138 OUS)
Distribution
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to c
Date initiated
2025-12-16
Date posted
2026-02-04
Location
Saint Charles, MO

Monitor recalls programmatically