Vortex Surgical 25GA I.D.D. Internal Delivery Device, VS0250.25
FDA device recall Z-1250-2026 · posted 2026-02-04 · Open, Classified
Recall details
- Recalling firm
- Vortex Surgical Inc.
- Reason for recall
- There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- Action taken
- On December 16, 2025, the firm began notifying customers via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were instructed to discontinue use of affected product and to return product for Vortex Surgical to evaluate the seals; use UPS account 23623Y to ship unused products to Vortex Surgical Inc for replacement or refund at: Attn: Andrew Ritts Recall Tyvek Seal, Vortex Surgical Inc., 4 Research Park Dr., Suite 124, St. Charles, MO 63304; complete and return the Acknowledgement and Receipt Form via mail or email to: ARitts@VortexSurgical.com. Actions have been implemented to improve training of individuals associated with sealing and inspecting seals to prevent reoccurrence. The sealer associated with the issue is being updated to include a guard to prevent reoccurrence. If you have any questions, contact the Director of Regulatory Affairs at 635-778-4350 or email to: ARitts@VortexSurgical.com from 8am to 5pm.
- Root cause
- Employee error
- Status
- Open, Classified
- Product code
- MXY
- Quantity affected
- 14,789 (8651 US; 6138 OUS)
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to c
- Date initiated
- 2025-12-16
- Date posted
- 2026-02-04
- Location
- Saint Charles, MO