FDA device recall Z-1177-2016 · posted 2016-03-15 · Terminated

Recall details

Recalling firm
Volk Optical Inc
Reason for recall
The firm discovered that the incorrect lot number was engraved on the product.
Action taken
Volk Optical sent an Urgent Medical Device Recall letter dated February 4, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately examine their inventory and quarantine any affected product. Also, if product was further distributed customers should identify their customers and notify them at once of the product recall and include a copy of the letter. The affected devices should be returned to Volk with an assigned Return Merchandise Authorization (RMA) number. Customers should contact Volk at (440) 942-6161 or toll free at (800) 345-8655 as soon as they receive the notification to setup a RMA for their device(s). Customers were instructed to complete and return the enclosed response form as soon as possible. Customers with questions should call (440) 510-0745.
Root cause
Employee error
Status
Terminated
Product code
HJK
Quantity affected
9 lens
Distribution
Worldwide Distribution - US (Virginia only) Internationally to Canada, France & Indonesia.
Date initiated
2016-02-02
Date posted
2016-03-15
Date terminated
2017-04-26
Location
Mentor, OH