Volcano SpinVision Sterile Equip cover II; Sterile; General Hospital: These Sterile Equipment covers are provided to cov

FDA device recall Z-1042-2016 · posted 2016-03-03 · Terminated

Recall details

Recalling firm
Volcano Corporation
Reason for recall
Volcano Corporation has identified that certain lot numbers of Volcano R-100 Sterile Equipment Covers may have the incorrect expiry date on the outer shipping box; individually wrapped inner product bears the correct expiry date.
Action taken
Urgent Medical Device letters were sent to affected consignees the week of February 8, 2016 by Fed Ex or UPS so delivery can be tracked. For OUS consignees, the same letter will be sent and they will execute delivery to customers and track responses.
Root cause
Labeling False and Misleading
Status
Terminated
Product code
KKX
Quantity affected
1355 total
Distribution
Distributed US (nationwide) and the countries of Australia, South Africa, Canada, New Zealand, Belgium, Japan, and Paraguay.
Date initiated
2016-02-05
Date posted
2016-03-03
Date terminated
2017-03-21
Location
Rancho Cordova, CA