Vivid S5 N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GY
FDA device recall Z-0754-2023 · posted 2022-12-22 · Open, Classified
Recall details
- Recalling firm
- GE Medical Systems, LLC
- Reason for recall
- Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire.
- Action taken
- GE Healthcare notified consignees on about 11/10/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION. For U.S. customers, letters were sent via FedEx. Consignees were instructed to follow updated safety instructions provided in the appendix included with the notification letter, place the appendix with the product labeling, and replace the batter every 2 years or if the battery is not capable of powering the system for more than 30 minutes. In response, GE Healthcare requested consignees return the response form.
- Root cause
- Device Design
- Status
- Open, Classified
- Product code
- IYN
- Quantity affected
- 1,400 units
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM
- Date initiated
- 2022-11-10
- Date posted
- 2022-12-22
- Location
- Waukesha, WI