VITROS Valproic Acid (VALP) Reagent, REF/Catalog No. 6801710 --- Ortho- Clinical Diagnostics Rochester, NY USA -- Ortho-

FDA device recall Z-0984-2012 · posted 2012-02-08 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Results using VITROS Chemistry Products VALP Reagent were negatively biased compared to other manufacturers. OCD confirmed that patient results were also affected by this issue.
Action taken
All domestic end-user consignees were initially sent "Important Product Correction Notification" recall communications (dated 5/10/11) by Federal Express overnight mail on 5/10/11. The letters informed users of the issue, and advised them to discontinue using their VITROS VALP reagent until adjusted calibrator values are provided for GEN 14 of VITROS VALP Reagent. The adjusted calibrator values for VITROS VALP Reagent were sent to customers via Assay Data Disk (ADD) Data Release Version (DRV) 5632. Customers were to refer to the Technical Support Communication for details. Users were to confirm receipt of the notice by completing and returning the Fax Reply Form - Confirmation; confirm that all customers or distributors were informed of the issue and instructed to discontinue use of affected product; and lastly confirm that all customers or distributors were sent adjusted calibrator values.
Root cause
Manufacturing material removal
Status
Terminated
Product code
LEG
Quantity affected
Approximately 5800 units worldwide (domestic - 4549, foreign - 1287)
Distribution
Worldwide Distribution -- US, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
Date initiated
2011-05-10
Date posted
2012-02-08
Date terminated
2016-02-12
Location
Rochester, NY