Atlas FDA

VITROS VALP Reagent , REF 680 1710. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests. Responsib

FDA device recall Z-0866-05 · posted 2005-06-04 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Negatively biased results may be observed when the reagent is stored on the VITROS 5,1FS Chemistry System for less than seven days.
Action taken
Letters dated 4/18/2005 with instructions to process a mid-level QC fluid every eight hours when processing patient samples for valproic acid. Replacement product to be shipped as soon as available.
Root cause
Other
Status
Terminated
Product code
LEG
Quantity affected
139 boxes
Distribution
37 U.S. consignees nationwide. Ortho-Clinical foreign affiliates in England, France and Spain.
Date initiated
2005-04-18
Date posted
2005-06-04
Date terminated
2005-10-13
Location
Rochester, NY