VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2
FDA device recall Z-2580-2024 · posted 2024-08-12 · Open, Classified
Recall details
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Reason for recall
- It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition codes which lead to a delay of results.
- Action taken
- On July 15, 2024, IMPORTANT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. REQUIRED ACTION 1. Send the enclosed Customer Letter and Confirmation of Receipt form to all customers who were shipped VITROS Immunodiagnostic Products Folate Reagent Pack 1/2 from your facility within the past 12 months. 2. Complete the enclosed Confirmation of Receipt form no later than July 23, 2024. If you have further questions, please contact the Global Services Organization at 1-800-421-3311.
- Root cause
- Software Design Change
- Status
- Open, Classified
- Product code
- CGN
- Quantity affected
- 18,467 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Nor
- Date initiated
- 2024-07-03
- Date posted
- 2024-08-12
- Location
- Rochester, NY
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| VITROS IMMUNODIAGNOSTIC PRODUCTS FOLATE REAGENT PACK 1/2, VITAMIN B12/FOLATE REAGENT PACK 3, PRODUCTS RED CELL FOLATE PA (K001266) | Ortho-Clinical Diagnostics, Inc. | 2000-05-02 |
| VITROS IMMUNODIAGNOSTIC PRODUCTS FOLATE REAGENT PACK 1/2, VITROS IMMUNODIAGNOSTIC PRODUCTS VITAMIN B12/FOLATE REAGENT P3 (K984166) | Ortho-Clinical Diagnostics, Inc. | 1999-02-05 |