Vitros Immunodiagnostic Products CEA REAGENT PACK, Lot 750, REF 192 0115, 100 coated wells per pack. Firm on the label:

FDA device recall Z-0299-04 · posted 2004-01-06 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.
Action taken
U.S. customers notified via FED EX letter sent on 12/3/03. Letter includes modified QC procedures to use until replacement product is issued. Foreign affiliates notified via e-mail on 12/3/03 with instructions to notify customers.
Root cause
Other
Status
Terminated
Product code
DHX
Quantity affected
524 reagent packs
Distribution
154 reagent packs shipped to consignees in the U.S. 370 reagent packs shipped to Ortho/J&J International Affiliates for further distribution to foreign users. Foreign consignees are located in Argenti
Date initiated
2003-12-03
Date posted
2004-01-06
Date terminated
2004-02-20
Location
Rochester, NY