VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin G antibody t

FDA device recall Z-2671-2014 · posted 2014-09-19 · Terminated

Recall details

Recalling firm
ORTHO-CLINICAL DIAGNOSTICS
Reason for recall
VITROS Immunodiagnostic Products Anti-HCV kit lot 9090 has been observed producing sporadic lower than expected VITROS Anti-HCV test results due to an approximate signal loss of up to 66.2% which can result in unexpected negative test results being obtained for reactive sample fluids (quality control and patient samples).
Action taken
All US consignees (users) who were shipped the affected lot VITROS HCV Reagent Pack were sent communications via FedEx and/or US Priority Mail (PO Boxes only) on 8/06/14, and advised of the issue and to immediately discontinue using all remaining inventory of VITROS Anti-HCV Reagent Packs Lot 9090. For additional information, please contact OCD's Customer Technical Service representatives at 1-800-421-3311.
Root cause
Packaging process control
Status
Terminated
Product code
MZO
Quantity affected
463 sales units
Distribution
NY, NJ, CT, MA, PA, VA, MD, FL and GA
Date initiated
2014-08-06
Date posted
2014-09-19
Date terminated
2017-04-14
Location
Bridgend, United Kingdom