VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin G antibody t
FDA device recall Z-2671-2014 · posted 2014-09-19 · Terminated
Recall details
- Recalling firm
- ORTHO-CLINICAL DIAGNOSTICS
- Reason for recall
- VITROS Immunodiagnostic Products Anti-HCV kit lot 9090 has been observed producing sporadic lower than expected VITROS Anti-HCV test results due to an approximate signal loss of up to 66.2% which can result in unexpected negative test results being obtained for reactive sample fluids (quality control and patient samples).
- Action taken
- All US consignees (users) who were shipped the affected lot VITROS HCV Reagent Pack were sent communications via FedEx and/or US Priority Mail (PO Boxes only) on 8/06/14, and advised of the issue and to immediately discontinue using all remaining inventory of VITROS Anti-HCV Reagent Packs Lot 9090. For additional information, please contact OCD's Customer Technical Service representatives at 1-800-421-3311.
- Root cause
- Packaging process control
- Status
- Terminated
- Product code
- MZO
- Quantity affected
- 463 sales units
- Distribution
- NY, NJ, CT, MA, PA, VA, MD, FL and GA
- Date initiated
- 2014-08-06
- Date posted
- 2014-09-19
- Date terminated
- 2017-04-14
- Location
- Bridgend, United Kingdom