VITROS Immunodiagnostic Products Anti-HBs Quantitative Reagent Pack, Catalog No. 6801925 VITROS Immunodiagnostic Product

FDA device recall Z-0772-2024 · posted 2024-01-19 · Open, Classified

Recall details

Recalling firm
Ortho-Clinical Diagnostics, Inc.
Reason for recall
Affected lots may experience increased calibration failures or an increase in falsely elevated results.
Action taken
On 27 November 2023, consignees who had been shipped the affected VITROS Anti-HBs Reagent Pack and Calibrator were notified of the issue via Urgent Production Correction Notification letters. Customers were instructed to immediately discontinue using, render unusable, and discard all affected lots of the affected product from their inventory. - QuidelOrtho will provide credit or replacement for the discarded product. - QuidelOrtho recommends a review of previous results between 12-30 mIU/mL, which were generated using an affected lot of VITROS anti-HBs Reagent Pack, where only anti-HBs testing was performed and was positive (for example, postvaccination testing) or where only anti-HBs was positive in the triple-panel test. If you have further questions, please contact QuidelOrtho's Global Services Organization (formerly Ortho Care) at 1-800-421-3311.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
LOM
Quantity affected
6,920 US; 57 OUS
Distribution
Domestic distribution nationwide. OUS distribution to Bermuda, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Mexico, Singapore, UK, France, Germany, Italy, Spain, Portugal, Russia,
Date initiated
2023-11-27
Date posted
2024-01-19
Location
Rochester, NY