Atlas FDA

VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemi

FDA device recall Z-1263-2015 · posted 2015-03-11 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Programming option; all other sample programming options function as intended.
Action taken
On 12/29/2014, Customer letter (Ref. CL2014-315) was sent via FedEx overnight courier or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to US consignees (OCD Direct, Drop-Ship and US Federal Government) to inform them of the issue and provided with instructions for reconfiguring their Barcode Scanners.
Root cause
Software design
Status
Terminated
Product code
N/A
Quantity affected
Domestic: 62 units; Foreign: 10 units
Distribution
Worldwide Distribution-US (nationwide) and the countries of Canada, Australia and England.
Date initiated
2014-12-29
Date posted
2015-03-11
Date terminated
2017-03-27
Location
Rochester, NY