Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Fir

FDA device recall Z-0387-05 · posted 2005-01-08 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Results may be negatively biased due to a non-linear response at the upper end of the Reportable (Dynamic) Range.
Action taken
Letters dated 11/22/2004 issued to customers. End users instructed to manually modify the Reportable and/or Analyzer Ranges; dilute all samples with an initial CRP result greater than 9.0 mg/dL. Distributors instructed to destroy product in-house and to notify end-users to manually modify the Reportable and/or Analyzer Ranges.
Root cause
Other
Status
Terminated
Product code
DCK
Quantity affected
Cat. 1926740 - 20,903 boxes; Cat 8097990 - 34,877 boxes
Distribution
Nationwide and worldwide distribution. 70% of product was shipped to foreign consignees.
Date initiated
2004-11-22
Date posted
2005-01-08
Date terminated
2005-05-02
Location
Rochester, NY