Atlas FDA

VITROS CHEMISTRY PRODUCTS TRIG Slides, REF 133 6544, IVD For in vitro diagnostic use only. VITROS Chemistry Products TRI

FDA device recall Z-0848-2015 · posted 2014-12-22 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Potential for biased results using the specific lot of VITROS CHEMISTRY PRODUCTS TRIG slides.
Action taken
Ortho Clinical Diagnostics sent an Urgent Product Correction Notification dated November 6, 2014, via FedEx overnight courierand/or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to all affected customers. The letter identified the product, the problem, and the acton to be taken by the customer. The letter informed the customers of the issue and advise them to discontinue use of the affected lot and to inform their Laboratory Medical Director if they suspect that a previously reported result may have been affected. For additional questions, please contact OCD's Customer Technical Services at 1-800-421-3311. For questions regarding this recall call 908-218-8776.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
CDT
Quantity affected
USA: 380 units, Foreign: 155 units
Distribution
Worldwide Distribution - USA (nationwide) Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore
Date initiated
2014-11-06
Date posted
2014-12-22
Date terminated
2017-02-28
Location
Rochester, NY

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
KODAK EKTACHEM CLIN. CHEM. SLIDES (TRIG) (K812029)Eastman Kodak Company1981-08-03