VITROS Chemistry Products LIPA Slides GEN 48, Coating 3235, REF/Catalog No. 166 8409 (60 slides per cartridge and 5 cart

FDA device recall Z-0245-2007 · posted 2006-12-05 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
The firm identified circumstances in which biased results or calibration failures can be observed when using any lot of VITROS LIPA slides if they have been stored on the VITROS Chemistry Systems for more than 2 days.
Action taken
On 10/09/06, Ortho-Clinical Diagnostics sent "IMPORTANT DISTRIBUTOR NOTIFICATION" Letters and "Confirmation of Receipt" Forms via Federal Express to all distributor accounts informing them of the "field correction", and instructing them to inform their customers who received these slides of this issue. OCD foreign affiliates were informed of this action by e-mail on 10/09/06, and instructed them to inform their customers who received the LIPA slides of this "field correction". OCD also sent "IMPORTANT PRODUCT CORRECTION NOTIFICATION" Letters and "Confirmation of Receipt" Forms to US end-user consignees that received the LIPA slides directly from OCD via US Postal Service, informing them of this "field correction" and that they may continue to use their current lot of Coating 3235 LIPA Slides until replacement product arrived providing that they followed the instructions in the letter.
Root cause
Other
Status
Terminated
Product code
CHI
Quantity affected
Cat./REF # 166 8409: Domestic - 3031, Foreign - 2511; Cat./ REF # 829 7749: Domestic - 1682, Foreign - 3553
Distribution
Worldwide distribution
Date initiated
2006-10-09
Date posted
2006-12-05
Date terminated
2008-05-22
Location
Rochester, NY

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
KODAK EKTACHEM CLINICAL CHEMISTRY SLIDES (K845027)Eastman Kodak Company1985-01-18