Atlas FDA

VITROS Chemistry Products-HCY Reagent packaged in a kit carton. Catalog# 6802002. Used to quantitatively measure total h

FDA device recall Z-1556-2013 · posted 2013-06-19 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
The firm is recalling the specific lot due to lower than expected results.
Action taken
Ortho Clinical Diagnostics sent an "Important Product Correction Notification" letter dated April 1, 2013 to all affected customers. Ortho Clinical Diagnostics recommends that their customers immediately discontinue using and discard all VITROS HCY Reagent remaining in your inventory for the affected lot. They have requested that their customers complete and return the attached Confirmation of Receipt form no later than April 12, 2013. Ortho Clinical Diagnostics requested that these notifications be forwarded to all laboratories and posted beside each VITROS System that utilizes VITROS HCY Reagent within their customers facilities. To place a product order for an alternate lot, please contact Ortho Clinical Diagnostics customer service representatives at 1-800-828-6316. The VITROS Chemistry Products HCY Reagent is used on the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System to quantitatively measure total homocysteine concentration in human serum and plasma.
Root cause
Process control
Status
Terminated
Product code
LPS
Quantity affected
478 total (339- Domestic) (139- International)
Distribution
Worldwide Distribution-USA (nationwide) including the states of AR, AZ, CA, CO, CT, FL, HI, ID, IL, KY, LA, ME, NJ, OK, TN, TX, WA, and WV, and the countries of Colombia, India, Japan, Mexico, Puerto
Date initiated
2013-04-01
Date posted
2013-06-19
Date terminated
2017-03-22
Location
Rochester, NY

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
VITROS CHEMISTRY PRODUCTS HCY REAGENT, VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 27, HCY PERFORMANCE VERIFIERS (K061588)Ortho-Clinical Diagnostics, Inc.2006-07-21