VITROS Chemistry Products GLU DT Slides GEN 56, REF (Catalog #) 153 2316, 25 slides per box. VITROS Chemistry Products G
FDA device recall Z-0151-2007 · posted 2006-11-04 · Terminated
Recall details
- Recalling firm
- Ortho-Clinical Diagnostics
- Reason for recall
- Positively biased results when using the VITROS Chemistry Products GLU DT Slides GEN 56.
- Action taken
- On 9/26/06, the recalling firm notified its foreign affiliate consignees of this 'field correction', and instructed them to inform their customers who received these slides of this issue, provided them with an updated Calibration Data Module and instructed to re-calibrate these slides for continued use. On 9/27/06, all US end-use consignees were informed of this issue via Federal Express overnight letter (dated 9/27/06), and instructing them to re-calibrate these slides for continued use after they have installed the Calibration Data Module. The new CDMs were sent to all customers with VITROS DT 60/DT60 II Chemistry systems starting 9/27/06 concurrent with, but separate, from the correction letters. This includes consignees who purchased their GLU DT slides from distributors. On 9/28/06, all distributors were notified of this 'field correction', providing instructions and 'Questions & Answers' for them, and instructing them to inform their customers who received these slides of this issue. Calibration Data Modules were sent to all customers with VITROS DT 60/DT60 II Chemistry Systems.
- Root cause
- Other
- Status
- Terminated
- Product code
- CGA
- Quantity affected
- 93,252 boxes (25 slides per box) in total
- Distribution
- Worldwide Distribution
- Date initiated
- 2006-09-27
- Date posted
- 2006-11-04
- Date terminated
- 2012-08-03
- Location
- Rochester, NY