VITROS Chemistry Products ECO2 Slides packaged as 300 slides/pack catalog number 8262396 VITROS ECO2 Slides quantitative
FDA device recall Z-0312-2014 · posted 2013-11-15 · Terminated
Recall details
- Recalling firm
- Ortho-Clinical Diagnostics
- Reason for recall
- Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
- Action taken
- Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides. VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue. OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs. OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP successfully to resolve the slide dispense issue. In addition to the Confirmation of Receipt form, OCD also enclosed three (3) copies of a Request for Credit form for you to complete when you require credit for affected cartridges. Complete and return the Confirmation of Receipt by June 21, 2013. If you experience slide dispense condition codes more than once a week, call our Customer Technical Support at 1-800-421-3311. For questions regarding this recall call 585-453-4224.
- Root cause
- Equipment maintenance
- Status
- Terminated
- Product code
- KHS
- Quantity affected
- 205,534 units (137,887 units domestically & 67,647 units internationally)
- Distribution
- Worldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR,
- Date initiated
- 2013-06-12
- Date posted
- 2013-11-15
- Date terminated
- 2017-04-11
- Location
- Rochester, NY
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| VITROS CHEMISTRY PRODUCTS ECO2 SLIDES (K120765) | Ortho-Clinical Diagnostics, Inc. | 2012-10-24 |
| VITRO CHEMISTRY PRODUCTS ECO2 SLIDE & CALIBRATOR KIT 2 (K001133) | Ortho-Clinical Diagnostics, Inc. | 2000-04-24 |