VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry Products ECO2 Slides quantitati

FDA device recall Z-0510-2024 · posted 2023-12-07 · Open, Classified

Recall details

Recalling firm
Ortho-Clinical Diagnostics, INc.
Reason for recall
Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not function as intended. These affected slides (potentially intermittent, consecutive slides within a cart or across consecutive carts) have an abnormality which may cause positively or negatively biased total carbon dioxide (CO2) results leading to either overdiagnosis of acidosis or missing alkalosis, both which may prompt unnecessary therapies such as IV fluids and/or bicarbonate or overdiagnosis of alkalosis which may also result in unnecessary use of IV fluids and electrolyte replacement or failure to detect acidosis and necessary treatment.
Action taken
Ortho Clinical Diagnostics (QuidelOrtho) issued Urgent Product Correction Notification letter to Distributors and End-Users on 10.25.23. Letter states reason for recall, health risk and action to take: Discontinue using, render unusable, and discard any remaining inventory of the affected lots of VITROS ECO2 Slides from your laboratory. " If available, use an alternate, unaffected lot of VITROS ECO2 Slides for testing. " Complete and return the enclosed Confirmation of Receipt form no later than November 2, 2023. QuidelOrtho will send a replacement lot or credit your account. " Ensure your laboratory follows the calibration instructions provided in the VITROS ECO2 Slides Instructions For Use. " Please forward this notification if the affected product was distributed outside of your facility. " Save this notification with your User Documentation or post this notification by each VITROS 250/350/5,1 FS/4600/5600/XT 3400/XT 7600 System until your laboratory receives a lot of VITROS ECO2 Slides unaffected by this issue. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization (formerly Ortho Care.) We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-421-3311. Enclosure: Confirmation of Receipt Form (Ref. CL2023-248_CofR)
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
KHS
Quantity affected
622 units 300 Slides (5 cartridge containing 60 slides)
Distribution
Nationwide Foreign: Bermuda Canada
Date initiated
2023-10-25
Date posted
2023-12-07
Location
Rochester, NY