VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration

FDA device recall Z-0535-2025 · posted 2024-11-25 · Open, Classified

Recall details

Recalling firm
Ortho-Clinical Diagnostics, Inc.
Reason for recall
May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.
Action taken
QuidelOrtho issued Urgent Product Correction Notification letter on 10.21.24. Letter states reason for recall, health risk and action to take: " Discontinue using, render unusable, and discard all remaining inventory of lots manufactured from the affected coating(s) of VITROS Chemistry Products CRP Slides. " Complete and return the enclosed Confirmation of Receipt form. Upon receipt of your completed Confirmation of Receipt form, QuidelOrtho will credit or replace your discarded inventory. " Save this notification with your User Documentation or post this notification by each VITROS 250/350/5,1 FS/4600/5600/XT 3400/XT 7600 System in your laboratory until the issue has been resolved. " Please forward this notification if the affected product was distributed outside of your facility. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. Root Cause Our investigation is ongoing; however, additional testing has been implemented in our release process to mitigate this issue prior to the release of future coatings of VITROS CRP Slides
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
DCK
Quantity affected
7717 units
Distribution
Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS, MN, MT, NY, PA, UT and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 0B8, Chile, China, Colombia, Denmark
Date initiated
2024-10-11
Date posted
2024-11-25
Location
Rochester, NY

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
VITROS Chemistry Products CRP Slides; VITRO Chemistry Products Calibrator Kit 7 (K152433)Ortho Clinical Diagnostics2015-09-24
EKTACHEM CLINICAL CHEMISTRY SLIDE C-REACTIVE PROTEIN (CRP) (K953197)Johnson & Johnson Clinical Diagnostics, Inc.1995-07-31