Atlas FDA

VITROS Chemistry Products CDM PROM Calibration Data Module Rev. No. 0168 and 0169 (Product Code 1999077) are used in con

FDA device recall Z-2214-2014 · posted 2014-08-13 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Inability to calibrate or use VITROS NH3 DT Slides, GEN 59 with VITROS Chemistry Products DT Calibrator Kit, Lot 44 when CDM PROM 0168 or 0169 is installed on VITROS DT60/DT60 II Chemistry Systems.
Action taken
Ortho Clinical Diagnostics sent an Urgent Product Correction Notification letter dated July 15, 2014,by Federal Express overnight to all affected customers. All US end-user consignees who had been shipped VITROS NH3 DT slides, GEN 59, Product Code 1532589 were informed of the issue. Customers were instructed to: a) use VITROS DT NH3 slides, GEN 59 with Calibration Data Module, Rev. 0168 or 0169 with VITROS DT Calibrator, Lot 42 if available; or b) If VITROS DT Calibrator, lot 42 is not available, the user is instructed to request CDM, Rev. 0170 to be used with VITROS NH3 DT slides, GEN 59 and VITROS DT Calibrator, Lot 44. Customers were instructed to forward this notice if the product was further distributed outside of their facility. OCD has manufactured a new CDM, Rev. 170, containing the necessary data/calibration mathematics for VITROS DT Calibrator, Lot 44 to be used to calibrate VITROS NH3 DT slides, GEN 59. Customers with questions were instructed to call Customers Technical Services at 1-800-421-3311. For questions regarding this recall call 908-218-8776.
Root cause
Process control
Status
Terminated
Product code
JIX
Quantity affected
5906 total: 240 (US); 5666 (OUS)
Distribution
Worldwide Distribution - USA including MA, TX, KY, IL, MS, OH and Puerto Rico; Internationally to: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mex
Date initiated
2014-07-15
Date posted
2014-08-13
Date terminated
2018-05-21
Location
Rochester, NY