VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Cod
FDA device recall Z-0220-2023 · posted 2022-11-10 · Open, Classified
Recall details
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Reason for recall
- Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results.
- Action taken
- Ortho Clinical Diagnostics issued Urgent Letter (CL2022-260) to VITROS 250/350 Chemistry System consignees. Letter states reason for recall, health risk and action to take: Upon receiving this communication, load Calibration Diskette DRV 6201 on your VITROS 250/350 Chemistry System and calibrate VITROS PHYT Slides, GEN 22 using VITROS Calibrator Kit 9. Complete the enclosed Confirmation of Receipt form no later than October 21, 2022. Please forward this notification if the affected product was distributed outside of your facility. Save this notification with your user documentation or post this notification by each VITROS 250/350 Chemistry System until Calibration Diskette DRV 6201 has been received and loaded by your facility. If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Ortho Care" Technical Solutions Center. If you have further questions, please contact the Ortho Care Technical Solutions Center at 1-800-421-3311
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- DKB
- Quantity affected
- 17,396 Units
- Distribution
- Worldwide Distribution: US (Nationwide) and Countries (Foreign) including: Australia Bermuda DV04 Bermuda, HM 13 Bermuda, HM08 Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India I
- Date initiated
- 2022-10-13
- Date posted
- 2022-11-10
- Location
- Raritan, NJ