VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- U

FDA device recall Z-2129-2023 · posted 2023-07-11 · Open, Classified

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration.
Action taken
QuidelOrtho (formerly Ortho Clinical Diagnostics) issued Urgent product correction Notification on 5/8/23. Letter states reason for recall, health risk and action to take: " Immediately discontinue using and discard your remaining inventory of VITROS Calibrator Kit 20, Lot 2022. QuidelOrtho will replace or credit your account. Indicate quantities to be replaced or credited on the attached Confirmation of Receipt form. " If you do not have an alternative lot of VITROS Calibrator Kit 20, you must discontinue reporting patient results using a Lot 2022 calibration. " Upon receipt of your replacement lot of VITROS Calibrator Kit 20, re-calibrate VITROS Transferrin, C3, C4, IgA, IgG and/or IgM assays on your VITROS System. " Complete the enclosed Confirmation of Receipt form . " Please forward this notification if the affected product was distributed outside of your facility. " Post this notification by your VITROS System(s) or with the user documentation until you receive your replacement order. Resolution Once we receive your completed Confirmation of Receipt form, QuidelOrtho will process your replacement order or credit your account. To provide replacement product for all customers, product allocation may be necessary. Contact Information . If you have further questions, please contact our Ortho Care Technical Solutions Center
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
JIT
Quantity affected
2400 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway,
Date initiated
2023-05-08
Date posted
2023-07-11
Location
Rochester, NY