VITROS Chemistry Products Calibrator Kit 16, REF 680 1700, 5x1 mL --- Ortho-Clinical Diagnostics, Rochester, NY USA -- O

FDA device recall Z-0714-2012 · posted 2012-01-11 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Ortho Clinical Diagnostics is recalling Lot 1630 of VITROS Calibrator Kit 16 because it will not likely meet its specified 12 months of stability.
Action taken
A letter was sent to customers via overnight delivery on 11/10/2011. The letter informed customers they should discontinue using and discard all remaining VITROS Calibrator Kit 16, Lot 1630; Complete and return the attached Confirmation of Receipt form upon receipt of the notification. Once the fax reply is received, the firm will begin the replacement order process; forward the notification to all departments and areas that use the VITROS Calibrator Kit 16, Lot 1630; and post the notification by each VITROS Chemistry or Integrated System in your facility that uses the VITROS Calibrator Kit 16, Lot 1630 or with VITROS user documentation.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
DHR
Quantity affected
768 kits
Distribution
Class II Recall: Worldwide Distribution - USA (Nationwide) and the countries of Brazil, Canada, Chile, China, Colombia, England, France, Mexico, Panama, Singapore, Spain, Venezuela.
Date initiated
2011-11-10
Date posted
2012-01-11
Date terminated
2018-07-02
Location
Rochester, NY