Atlas FDA

VITROS Chemistry Products Calibrator Kit 11 VANC Reagent Generation (GEN) 50 with affected Assay Data Disk (ADD) for Dat

FDA device recall Z-1664-2023 · posted 2023-05-30 · Open, Classified

Recall details

Recalling firm
Ortho-Clinical Diagnostics, Inc.
Reason for recall
Unavailable to calibrate VITROS Chemistry Products VANC Reagent, Generation (GEN) 50, results in delay of testing
Action taken
QuidelOrtho (formerly Ortho Clinical Diagnostics) issued Urgent Product Correction Notification Letter on 4/14/23 to Distributors and End/Users. Letter states reason for recall, health risk and action to take: Prior to calibrating VITROS VANC Reagent GEN 50 using VITROS Calibrator Kit 11, Lot 1172, install ADD DRV 6224 or above on your VITROS System(s). " Complete the enclosed Confirmation of Receipt form no later than April 24, 2023. " Forward this notification if VITROS Calibrator Kit 11, Lot 1172 was distributed outside of your facility. " Save this notification with your user documentation or post this notification by each VITROS 5,1 FS/4600/5600/XT 7600 System until ADD DRV 6224 or above has installed. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Ortho Care" Technical Solutions Center. Contact Information If you have further questions, please contact the Ortho Care Technical Solutions Center at 1-800-421-3311.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
DLJ
Quantity affected
132 units
Distribution
Nationwide Foreign: Brazil, Chile, Colombia
Date initiated
2023-04-14
Date posted
2023-05-30
Location
Raritan, NJ