Atlas FDA

VITROS Chemistry Products BuBc Slides packaged as 300 slides/pack catalog number 8383051 and packaged as 90 slides/pack

FDA device recall Z-0300-2014 · posted 2013-11-15 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
Action taken
Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides. VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue. OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs. OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP successfully to resolve the slide dispense issue. In addition to the Confirmation of Receipt form, OCD also enclosed three (3) copies of a Request for Credit form for you to complete when you require credit for affected cartridges. Complete and return the Confirmation of Receipt by June 21, 2013. If you experience slide dispense condition codes more than once a week, call our Customer Technical Support at 1-800-421-3311. For questions regarding this recall call 585-453-4224.
Root cause
Equipment maintenance
Status
Terminated
Product code
CIG
Quantity affected
74,792 units (35,501 units domestically & 60,291 units internationally)
Distribution
Worldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR,
Date initiated
2013-06-12
Date posted
2013-11-15
Date terminated
2017-04-11
Location
Rochester, NY

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
EKTACHEM CLINICAL CHEM. SLIDES (K840606)Eastman Kodak Company1984-04-04