Vitros Chemistry *** GLU (glucose) DT Slides ***25 slides per box *** CAT 153 2316 *** Ortho-Clinical Diagnostics Inc. ,

FDA device recall Z-1225-03 · posted 2003-09-11 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Results of glucose measurements on sodium fluoride/potassium oxalate specimens can be negatively biased when using Vitros GLU and GLU DT slides stored in the refrigerator.
Action taken
GLU Slides: Letters dated between 7/28/03 and 8/11/03 provide revised storage instructions. Customers using NaF/KOx blood collection tubes and have VITROS GLU slides that have been refrigerated for more than four months will receive replacement product. Those customers are instructed to continue to use the slides only until replacement product arrives. GLU DT Slides: Letters dated between 7/28/03 and 8/11/03 with the same instructions as for GLU slides.
Root cause
Other
Status
Terminated
Product code
CGA
Quantity affected
210,050 boxes
Distribution
Nationwide and worldwide distribution to distributors and end-users.
Date initiated
2003-07-28
Date posted
2003-09-11
Date terminated
2003-11-24
Location
Rochester, NY