VITROS 5600 Integrated System, System Product Code 6802413. For use in the in vitro quantitative, semi-quantitative, and

FDA device recall Z-0078-2015 · posted 2014-10-16 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator barcode labels supplied with VITROS Chemistry Products Calibrator Kits 1, 2, 3, 4, 6, or 9, an unexpected assay calibration may occur if assay targets are unassigned (i.e., hidden).
Action taken
On 9/08/14, URGENT PRODUCT CORRECTION NOTIFICATION Letter (CL2014-232, dated 9/8/14) and Confirmation of Receipt Form were sent via FedEx overnight courier or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to US consignees (OCD Direct, Drop-Ship and US Federal Government only) to inform them of the issue and advise them of the required actions. Foreign affiliates were informed of the issue via e-mail on 9/8/14 and instructed to notify their consignees of the issue and the required actions.
Root cause
Software design
Status
Terminated
Product code
JJE
Quantity affected
USA: 843; Foreign: 945
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Sin
Date initiated
2014-09-08
Date posted
2014-10-16
Date terminated
2018-05-21
Location
Rochester, NY