VITROS 3600 Immunodiagnostic System --- For use in the in vitro quantitative, semi-quantitative and qualitative measurem
FDA device recall Z-2139-2014 · posted 2014-08-01 · Terminated
Recall details
- Recalling firm
- Ortho-Clinical Diagnostics
- Reason for recall
- Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.
- Action taken
- On 7/07/2014, Customer Letter (Ref. CL2014-191) was sent via FedEx overnight courier or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to US consignees (OCD Direct, Drop-Ship and US Federal Government) to inform them of the issue and advise them of the required actions. On 7/07/2014, Distributor Letter (Ref. DL2014-191 was sent via US Postal Service Priority Mail and/or FedEx overnight courier to US Distributor consignees instructing them to notify their consignees of the issue and required actions. Foreign affiliates were informed by e-mail on 7/07/2014 of the issue and instructed them to notify their consignees of the issue and required actions. For questions, please contact Customer Technical Services at 1-800-421-3311.
- Root cause
- Software in the Use Environment
- Status
- Terminated
- Product code
- JJE
- Quantity affected
- Domestic: 142, Foreign: 536
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Sin
- Date initiated
- 2014-07-07
- Date posted
- 2014-08-01
- Date terminated
- 2018-05-21
- Location
- Rochester, NY
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| VITROS IMMUNODIAGNOSTIC PRODUCTS TSH REAGENT PACK AND CALIBRATORS, VITROS 3600 IMMUNODIAGNOSTIC SYSTEM (K083173) | Ortho-Clinical Diagnostics | 2008-12-22 |