VITROS 250 Chemistry Systems, Catalog 8132086, Unique Device Identifier No. 10758750004409, and VITROS 250 Refurbished,

FDA device recall Z-1071-2016 · posted 2016-03-08 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).
Action taken
Ortho Clinical Diagnostics sent an URGENT PRODUCT CORRECTION NOTIFICATION Letter (Ref. CL2015-201, dated December 21, 2015) via FedEx overnight courier and/or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address). The notification letter was sent to all customers who received the affected product. Foreign affiliates were informed by e-mail on 12/20/2015 of the issue and instructed to notify their consignees of the issue and required actions. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Technical Services at 1-800-421-3311.
Root cause
Error in labeling
Status
Terminated
Product code
JJE
Quantity affected
US: 802; Foreign: 2472
Distribution
Worldwide Distribution - US Nationwide including Puerto Rico, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Sin
Date initiated
2015-12-21
Date posted
2016-03-08
Date terminated
2018-05-30
Location
Rochester, NY