Vitatron C-Series C60DR DDDR Dual Chamber Rate Responsive Pacemaker includes the following models: C60A1 and C60A3. The

FDA device recall Z-0999-2007 · posted 2007-07-07 · Terminated

Recall details

Recalling firm
Medtronic Inc. Cardiac Rhythm Managment
Reason for recall
Pacing rate; Software anomaly affecting Vitatron dual chamber C-series and T-series pacemakers, related to automatic retrograde conduction testing, can inhibit the device from pacing if the patient's intrinsic rate falls below the programmed lower rate.
Action taken
Advisory letters to physicians were mailed in February 2007. Sales respresentatives will follow-up with physicians by visit. A programmer software update will be released upon regulatory approval. The Advisory Letter provided short-term direction to reduce problems pending software upgrades.
Root cause
Software design
Status
Terminated
Product code
NVZ
Quantity affected
4,294
Distribution
within the US to include: AL, AR, AZ, CA, CT, FL, GA, IL, IN, KY, MA, MD, MI, MN, NC, ND, NJ, NY, OH, OK, SC, TX.
Date initiated
2007-02-05
Date posted
2007-07-07
Date terminated
2008-10-16
Location
Minneapolis, MN